JT-Series High and Low Temperature Test Chamber
Product Features
The whole system is equipped with 10 inch color capacitive touch screen as standard, which provides better user experience and improves the convenience of touch operation.
Using independent intelligent control logic system, can be more accurate control, with a wider temperature range to meet the user's diverse testing needs.
The inner liner is made of SUS304 mirror stainless steel, with excellent corrosion resistance, easy to clean the surface, and more convenient maintenance.
Advanced energy-saving design, the use of intelligent control system + special air circulation system can be realized automatically adjust the operating parameters of the equipment, to achieve the effective use of energy.
JT-Series High and Low Temperature Test Chamber | ||||||
Model | JT330(C) | JT610(C) | JT1000(C) | JT1530(C) | JT2000(C) | |
Nominal inside volume(L) | 335 | 605 | 1000 | 1530 | 2000 | |
Internal dimension(mm) | D | 760 | 795 | 960 | 1470 | 1670 |
W | 585 | 800 | 1100 | 1100 | 1100 | |
H | 750 | 950 | 950 | 950 | 1100 | |
External dimension(mm) | D | 1820 | 1850 | 2000 | 2525 | 2700 |
W | 880 | 1100 | 1450 | 1450 | 1430 | |
H | 1950 | 2230 | 2230 | 2230 | 2400 | |
Performance Parameters | Minimum temperature ºC | End with C: minimum temperature -70ºC; without C: minimum temperature -40ºC | ||||
Maximum temperature ºC | +180ºC(Maximum temperature can be up to 220ºC according to user's actual demand) | |||||
Temperature fluctuation ºC | ±0.1ºC~±0.5ºC | |||||
Temperature uniformity ºC | <=2ºC(≤150ºC) | |||||
Temperature deviation ºC | <=±2ºC | |||||
Temperature rise rate ºC/min | 4.0 | 6.0 | 4.0 | 4.0 | 3.0 | |
Cooling rate ºC/min | 3.5 | 2.6 | 2.8 | 2.5 | 2.0 | |
Test method | The above indexes are measured at no load, ambient temperature 20±2ºC (according to GB/T 5170.2-2017) Temperature and humidity performance test according to standard IEC60068-3-5, the sensor is placed at the air outlet of the air handling unit. |
|||||
Meets standards | GB2423.1,GB2423.2,GJB150.3,GJB150.4,GB11158,GB10589 | |||||
Thermal compensation capacity | 1500-3000 W | |||||
Cooling method | Air-cooled/water-cooled (optional) | |||||
Power supply mode | AC380 V ± 10% 50Hz 3/N/PE | |||||
Maximum power KW | 4.8 | 8.5 | 11.5 | 11.5 | 15.5 | |
Maximum current A | 12.3 | 14.8 | 22 | 22 | 30 | |
Noise level Db(A) | 65 | 65 | 68 | 68 | 70 | |
Weight KG | 480 | 615 | 870 | 1000 | 1250 |