Brand name: SARTEC
Machine name: Computer System VICAT/HDT Tester
Machine model:SS-3900HV
I , Purpose
This machine is suitable for testing plastic test piece VICAT/HDT . This
machine adopts computer system control, high precision LVDT special
displacement sensor. The deformation value is automatically locked after
reaching the standard temperature. Testing is simple and convenient. Test
temperature up to 300ºC, equipped with water cooling system, to achieve
rapid cooling.
II,Machine features
- Computer system control, simple operation, integrated control, automatic timing, real-time temperature display. Program automatic control.
- The initial load weight of the frame is light, which meets the requirements of the new standard for the minimum load of flat samples
- The sample frame, indenter and load rod are made of high- performance abrasive steel with high hardness and low coefficient of thermal expansion.
- Deformation induction device: use special LVDT displacement sensor, resolution is 0.001mm
- The load rod has an air guide ring design to provide vertical frictionless action
- Built-in heat exchanger, water cooling can be carried out. Improve test
efficiency.
- The software automatically corrects the coefficient of thermal expansion and eliminates the system error of themachine
- Adopt high precision temperature controller and temperature sensing rod to form closed-loop temperature control system. Temperature multi-point linear correction calibration function
- The method of circulating stirring with multiple impellers ensures the uniform temperature of the test oil. At the same time, three groups of temperature sensing rods are used to test the deformation temperature beside the sample respectively.
- Reach the standard deformation, automatically record the corresponding temperature value, and automatically shut down.
III, Main technical indicators
- According to the standard: GB/T1633, GB/T1634, etc
- Test mode: VICAT/HDT
- Control mode: computer system operation
- Sampling frequency: 33HZ
- Temperature control range: room temperature -300ºC
- Precision of temperature control: ±0.1ºC
- Temperature accuracy: ≤±1ºC
- Heating rate: 50ºC/h, temperature rise of 5ºC±0.5ºC every 6min 120ºC/h, temperature rise of 12ºC±1ºC every 6min
- Deformation measurement: LVDT displacement sensor
- Deformation measurement range: 0.001~4mm
- Deformation measurement resolution: ±0.001mm
- Deformation measurement accuracy: 0.1~1mm range ≤0.005mm
- Thermal deformation sample span: 64mm, 100mm other optional
- Temperature sensing: measurement around the sample.
- Number of test stations: 3 or 6 groups optional
- Heating medium: methyl silicone oil (below 200 CST, flash point above 300ºC, customer's own)
- Cooling mode: water cooling mode, improve the test efficiency.
- Volume (W ×D×H) : 1500mm×840mm×1710mm(W*D*H)
- Weight: about 355kg
-
- Power supply: AC220V 50HZ 25A
-
- Customer preparation:
- Installation space, flat bearing level ground
l Heat conduction oil: about 30L, flash point requirements of 300ºC or above. 100# recommended
- Power requirements: AC(220±10)V, (50±0.5)Hz, single phase two wire + protection ground, customer's own 32A, 2P circuit breaker.
- Ground cable: The ground resistance of the grounding cable is less than 4 ohms
- Water source: upper and lower water, water into tap water, water into the ground drain or sewer pipe can be, cooling for long water
IV, Machine configuration
- Host: 1 PC
- Test load Weights: 3 sets (or 6 sets)
- HDT indenter: 3 (or 6 sets)
- HDT supports: 3 (or 6 sets)
- VICAT indenter: 3 (or 6 sets)
- VICAT supports: 3 (or 6 sets)
- Instructions in Chinese: 1 copy
- Third party test report: 1 copy
List of random attachments: 1