Customization: | Available |
---|---|
After-sales Service: | Response Within 24 Hour |
Warranty: | 1 Year |
Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
---|
Payment Methods: |
|
---|---|
Support payments in USD |
Secure payments: | Every payment you make at Made-in-china.com is protected by the platform. |
---|
Refund policy: | Claim a refund if your order doesn't ship, is missing, or arrives with product issues. |
---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Type no. | ADV100 ADL100 |
ADV150 ADL150 |
ADV250 ADL250 |
ADV500 ADL500 |
ADV1000 ADL1000 |
Temperature range | Room temperature +10 C to 250 C | Room temperature +10ºC ~ 300ºC | |||
Maximum temperature arrival time | Approx.70min | ||||
Nominal content product | 80L | 150L | 300L | 500L | 1000L |
Inner box size mm Width W x depth D x height H |
400 * 400 * 500 | 600 * 500 * 500 | 600 * 500 * 1000 | 800 * 700 * 900 | 1000 * 1000 * 1000 |
External dimensions MM Width W x depth D x height H |
510 * 550 * 870 | 710 * 650 * 870 | 710 * 650 * 1610 | 910 * 860 * 1610 | 1200 * 1650 * 1500 |
Way to work | ADV: Program run/Value run ADL: Value run | ||||
The weight of the | About 60 kg | About 75 kg | About 120 kg | About 180 kg | About 300 kg |
The power capacity | AC220V 6.0 A | AC220V 7.5 A | AC220V 15A | AC380V 20A | AC380V 18A |
1. Product name and purpose | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
1.1 Product Name 1.2 use 1.3 Execution Standards |
Air supply drying oven with fixed temperature Suitable for electronic semiconductor, liquid crystal, precision instruments, materials manufacturing, life science and other aspects of baking, drying, curing and constant temperature test. GB/T 30435-2013 electric heating drying oven and electric heating blast drying oven |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
2. Circulation mode | Forced air circulation | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
3 Sample limit | Test or storage of flammable, explosive, volatile material samples corrosive material sample test or storage Testing or storage of biological samples Strong electromagnetic emission source test or storage of samples |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
4.Model naming method |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5. Volume, size and weight | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1 Standard volume of contents 5.2 Inner dimensions 5.3External Dimensions 5.4 weight |
See [Main Technical parameters] See [Main Technical Parameters] See [Main Technical Parameters] See [Main Technical Parameters] |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Performance of 6. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
6.1Test environmental conditions 6.2 Test Methods 6.3Temperature Range 6.4Temperature fluctuation 6.5 Temperature uniformity 6.6 Heating Time 6.7 Temperature Control |
The ambient temperature is 23 ° C ±5 ° C, the humidity is 65%±20%, and there is no sample in the device GB/T 30435-2013 electric heating drying oven and electric heating blast drying oven See [Main technical parameters] ±0.5 ° C (no-load test) ±2.5% (no-load test) See [Main Technical parameters] Use armoured SUS304 stainless steel heater as a heating element, intelligent temperature control instrument for real-time sampling box temperature, compared with the temperature on the controller set point, the PID arithmetic, to control the on-off time of SSR, so as to control the output of the heater heat, at the same time combined with forced air circulation way, achieve constant temperature in the oven. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
7. Structural features | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
7.1 Exterior materials 7.2 Interior materials 7.3 Heat break material 7.4 the door 7.5 the door seal 7.6 observation window 7.7 castor 7.8 vent 7.9 lead hole 7.10 Shed board load bearing 7.11 Number of shed support sections 7.11 Distance between shelter reception segments 7.12 the heater 7.13 the fan |
Cold rolled steel plate, surface electrostatic spray treatment, color RAL9003 signal white ADV series: SUS304 stainless steel plate; ADL series: SUS201 stainless steel plate; High density aluminum silicate fiber cotton Single door with hinge on left and handle on right Seal with Japanese-made custom foam silicone seal strips
30mm Stainless steel tube heater
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
8 Control system | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
8.1 temperature modulator 8.2 Temperature control mode 8.3 Temperature setting Mode 8.4 Temperature Setting Method 8.5 Operation Function 8.6 Program Mode 8.7 Accessory Features 8.8 Heater loop control 8.9 Temperature overrise preventer 8.10 the sensor 8.11 the power cord |
Imported intelligent temperature controller PID control Special operation setting menu and key for digital setting 4 section LED digital display ADV series: fixed value run, program run, ADL series: fixed value run 2 sets, 30 segments (ADV series only) Temperature compensation function, key lock function, PID self-tuning function SSR (solid state relay) contactless isoperiodic pulse width tuning It is used to prevent abnormal overrise of temperature Both the temperature regulator and the overrise preventer use K-type thermocouples Adv100/150, ADL100/150: about 2m long with plug ADV250//500/1000, ADL250/500/1000: about 3m long wiring |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
9. Safety device | Self-diagnosis function (temperature sensor abnormality, heater disconnect, SSR short circuit, automatic overrise prevention function), lock key function, overrise prevention device, leakage circuit breaker | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
10. Accessories | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
10.1 Instruction Manual 10.2 shed board |
Instructions for detailed use of the product
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
11 Optional accessories | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
11.1 stand 11.2 shed board 11.3 lead hole 11.4 External communication function 11.5 Overlapping shelf |
The bottom surface is fixed with ADV100/ADV150;No ADV250/500/1000 For when additional ADV250,500/1000 is required Inner diameter 25/50mm right side of body RS485 |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Transport of 12. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
12.1 way 12.2 Packing Materials |
Integral, integral transport Packed in wooden cases |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
13. Service conditions | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
13.1 site 13.2 Environmental conditions 13.3 Power capacity 13.4 Requirements for the storage environment |
The floor is level and well ventilated There is no strong vibration around the equipment There is no strong electromagnetic field around the equipment No flammable, explosive, corrosive substances, or dust are found around the device The ground bearing capacity: not less than 800kg/m2 Appropriate space for use and maintenance should be left around the equipment Temperature: 0ºC ~ 40ºC Relative humidity: ≤85% Air pressure: 86kPa ~ 106kPa See [Main technical parameters] When the equipment is not working, the ambient temperature should be kept within 0 ° C to +35 ° C |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
15.After-sales service | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
15.1 the quality assurance period 15.2 Out of warranty period 15.3 Timeliness of service |
Within 12 months from the date of acceptance and signature confirmation, the company shall be responsible for free maintenance of faults caused by manufacturing reasons, but not including equipment damage caused by human reasons; For equipment maintenance outside the warranty period, only the cost of accessories will be charged and lifelong technical support will be provided; ATEC company has a spare parts library, can quickly provide a variety of parts, outside the warranty period, only the cost charge.After the seller receives the buyer's fault notice, guarantee to respond within 1 hour, and strive for 48 hours to troubleshoot; |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
14. The appended drawings |